6.6 Remove and Install Pumps
The Vertex M system includes two vacuum pumps. One pump operates while the other is idle. A
defective pump may be replaced while the other pump continues to operate.
Note: A pump may be replaced only when the system places it in standby. Do not replace an
operating pump.
To equalize wear on the pumps, alternate the operating pump every 6 months.
To change operating pump without manually stopping monitoring touch the Main Screen,
Menu, Service, Authorized Service buttons, and then the Pump Alternate button.
The operating pump indicator is green or yellow; the indicator for the pump in idle is white.
6.6.1 Remove Pump
1. Open lower cabinet door and remove filter.
2. Turn thumbscrews counterclockwise to release and open pump enclosure door.
3. Push slide plate to disengage.