Honeywell Vertex M - H 2 S LL Hydrogen Sulfide (Low Level); 2 Nitrogen Dioxide; NF 3 Nitrogen Trifluoride; Phosgene

Honeywell Vertex M
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gas Name
1
TLV LAL LDL
Default
Alarm
Level 1
Default
Alarm
Level 2
Range Alarm Setting
Initial Ana-
lysis Period
(second)
Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft.
Sample Line)
Chemcassette
Part Number
Hydrogen Sulfide
(H
2
S) (Hydrides)
1 ppm 2 ppm 1 ppm 10 ppm 20 ppm 0-50 ppm
2-9.9 ppm
10-19.9 ppm
20-39.9 ppm
40-50 ppm
60
45
30
15
<20 sec (Alarm @ 10 ppm with 20 ppm H
2
S gas) 1295-0300
Hydrogen Sulfide
(H
2
S) Low Level
1 ppm 10 ppb 8 ppb 250 ppb 500 ppb 0-2000 ppb
10-99 ppb
100-499 ppb
500-999 ppb
1000-2000 ppb
480
240
120
60
<40 sec (Alarm @ 500 ppb with 1 ppm H
2
S gas) 1295-0223
Hydrogen Sulfide XP
(H
2
S)
1 ppm 1 ppm 0.5 ppm 10 ppm 20 ppm 0-50 ppm
1-4.9 ppm
5-9.9 ppm
10-50 ppm
30
15
10
<20 sec (Alarm @ 10 ppm w/ 20 ppm H
2
S gas) 1295-0226
Hydrogen Sulfide
XP4 (H
2
S)
1 ppm 1 ppm 0.5 ppm 10 ppm 20 ppm 0-50 ppm
1-4.9 ppm
5-9.9 ppm
10-50 ppm
30
15
10
<20 sec (Alarm @ 10 ppm with 20 ppm H
2
S gas) 1257-9300
Hydrogen Sulfide XP
(H
2
S) Low Level
1 ppm 20 ppb 15 ppb 500 ppb 1000 ppb 0-2000 ppb
20-99 ppb
100-199 ppb
200-399 ppb
400-2000 ppb
120
60
30
15
<25sec (Alarm @ 500 ppb with 1000 ppb H
2
S gas) 1295-0226
Hydrogen Sulfide
XP4 (H
2
S) Low Level
1 ppm 20 ppb 15 ppb 500 ppb 1000 ppb 0-2000 ppb
20-99 ppb
100-199 ppb
200-399 ppb
400-2000 ppb
120
60
30
15
<15sec (Alarm @ 500 ppb with 1000 ppb H
2
S gas) 1257-9300
Nitrogen Dioxide
(NO
2
)
0.2 ppm 0.5 ppm 0.3 ppm 3 ppm 6 ppm 0-30 ppm
0.5-8.9 ppm
9-30 ppm
240
120
<130 sec (Alarm @ 3ppm with 6ppm NO
2
gas) 1295-0220
Nitrogen Trifluoride
(NF
3
) Pyrolyzer Only
10 ppm 1 ppm 1 ppm 10ppm 20 ppm 0-50 ppm
1-9.9 ppm
10-19.9 ppm
20-39.9 ppm
40-50 ppm
40
20
10
10
<70 sec 1295-0225
Nitrogen Trifluoride
XP (NF
3
) Pyrolyzer
Only
10 ppm 1 ppm 1 ppm 10 ppm 20 ppm 0-50 ppm
1-9.9 ppm
10-19.9 ppm
20-39.9 ppm
40-50 ppm
30
15
10
10
<20 sec 1295-0507
Nitrogen Trifluoride
XP4 (NF
3
) Pyrolyzer
Only
10 ppm 1 ppm 1 ppm 10 ppm 20 ppm 0-50 ppm
1-9.9 ppm
10-19.9 ppm
20-39.9 ppm
40-50 ppm
30
15
10
10
<20 sec 1257-9310
Phosgene XP
(COCl
2
)
100 ppb 10 ppb 7 ppb 100 ppb 200 ppb 0-1000 ppb
10-49 ppb
50-99 ppb
100-199 ppb
200-1000 ppb
60
45
30
15
<25 sec 1295-0228
TLV - Threshold Limit Value LAL - Lowest Alarm Level LDL - Lower Detectable Limit
1. ā€œ-1ā€ calibrations allow more stain development but will extend response times
2. Minimum Sample Time