8.10.4 Configuration Requirements
Analog output cards should be assembled with an Ethernet IP Adaptor module that should be
connected to PLC processor module via Ethernet connection. The one end of RJ45 cable is
connected to the Link A2 of processor module. This is the rearmost connector, near to the DIN
rail and far from the LEDs. The other end connects to the Link A2 of the Ethernet IP Adapter
module as shown in the picture below. The IP address of the Ethernet IP adaptor should be
configured to 192.168.1.34. (pre-configured at factory)
8.10.5 Operational Description
The analog outputs will mirror the corresponding Fieldbus concentration register. See "Data
Values Common to Fieldbus Networks" on page198 for more information. This section fully
describes the outputs under normal operating conditions.
Condition Channel Output Notes
Normal
Concentration
Value or 2 mA fault
indicator
See "Data Values Common to Fieldbus Networks" on
page198 2ma fault indicator must be enabled and is off
by default.
Com
Failure
0 mA on all
channels
Communication link failure between PLC and Ethernet
Adaptor
PLC
Failure
0 mA on all
channels
Including PLC faulted or in program mode
Vertex M 264 User Manual