7.5 Information Events
The Vertex M system enters informational and other non-fault events into the event history database. These do not require any action by the user.
Use the event history to check the status of the monitor.
Event Type In History Event History Message Possible Cause
ALM SIMULATE Various An alarm was artificially created. See "Security Access" on page114 for more information.
ALM 1 Concentration > AL1 Gas concentration exceeded limit
ALM 2 Concentration > AL2 Gas concentration exceeded limit
ALM WARNING Concentration > Warning Gas concentration exceeded lower detection limit
AZ NO RECORD Various A fault was cleared in the DAq for which the analyzer has no record
AZ PROGRAM INSTALL Various The analyzer program update process was started
CALIBRATION Command – Valves Auto Balance Auto Balance was started
CONFIG INSTALL Various The configuration profile installation was started
DEFAULT RESET Various A fault was cleared in the DAq from an analyzer that is no longer present in the rack
FAULT SIMULATE Various A fault was artificially created. See "Security Access" on page114 for more information.
INFO 20 mA Loop Test Ended User initiated action
INFO 20 mA Loop Driven for Test User initiated action
INFO Window Zero Reset Optics defect detected and corrected
INFO Analyzer accepts new location Analyzer moved
INFO Analyzer Powered Up User initiated action
INFO Az Button Resets Alm&Flts Internal override pushbutton operated
INFO Az Button Starts Monitoring Internal override pushbutton operated
INFO Az Button Stops Monitoring Internal override pushbutton operated
INFO Clock Adjusted By DAq PC Normal drift or clock set on DAq
INFO Flow Auto Balance Performed User initiated action
INFO Hardware Configuration Changed Auto Balance, Load CC, New Location or Pump Alternated
INFO LED Drive Adjusted Load CC
INFO Monitoring stopped for no pts All points disabled by faults
INFO New Chemcassette Was Installed User initiated action
INFO New Sector Started Normal maintenance
INFO Optics Verified Successfully Load CC
INFO Program Loaded User initiated action
INFO Pump Selection Resynchronized Analyzer installed in different system
INFO Q-Factor Set Load CC
INFO Rebooted W Watchdog reboot of analyzer