Event
Number
Event List Desc.
(Short)
Event History
Description
Data Field Problem Possible Cause Recovery
235
AZ SW VER
WRONG
Analyzer software
version mismatch
0 Software mismatch
Reload program to all analyzers
Contact Honeywell service
237
AZ VACUUM
LOW
Inadequate
Analyzer Vacuum
Observed
vacuum in in
Hg
Individual analyzer observes
inadequate vacuum, stops
monitoring
Fault 219 has triggered. Recover from Fault 219, then reset Fault 237
Circular tubing harness disconnected Connect circular tubing harness
Tubing pinched
Correct tubing dress in cable carrier
Contact Honeywell Service
238
OP SW
MISMATCH
Optics block
software
mismatch.
0
The two optics blocks are not
running the same software
version.
Program update done incorrectly. Reload program to both optics blocks.
239
OPTICS COMM
FAIL
Optics block
communication
failure
Error code
Analyzer is not able to
communicate with the optics
block.
Cable disconnected.
Software anomaly.
Defective optics block.
Check the 14 pin cable connection on the sensor board.
Cycle power to analyzer.
Contact Honeywell service.
240
OPTICS REF
FAIL
Optics reference
limit exceeded
Reference
value signal
The signal detected at the
reference photodiode is out of
range.
Degraded optics block. Contact Honeywell service.
241
OPT REF ADJ
TIMEOUT
Optics reference
adjustment
timeout
Error code
Optics block self-adjustment
timeout
Cycle power to analyzer. Contact Honeywell service.
270
COMAZ ##
BROKEN
OPC To Az ##
Communications
Broken
Analyzer not responding to OPC
driver polling request.
Analyzer powered down Check switch on power module
Ethernet cable not connected
Check Ethernet connection at the back of the analyzer.
Move cable to different port on Ethernet hub.
301
PLC-COM
BROKEN
PLC-Com Broken
PLC fails to respond to DAq
communication
Cable connection problem
Series 1 Vertex M PLC
Check serial communications on DAq COM2
Check serial cable on PLC 1747-KE (See "Data Acquisition Computer
(rear)" on page19 for more information.) (See "Main PLC" on page22
for more information.)
Series 3 Vertex M PLC
Check Ethernet connection between PLC and network switch
Loss of PLC power Check power connections for PLC
302
AZ-BAD PROF
ID
AZ has wrong
profile loaded
Profile ID in analyzer does not
match current configuration in
use
Configuration not properly loaded Reinstall current Configuration Profile
303
PLC-BAD PROF
ID
PLC Has Wrong
Profile loaded
Profile ID in PLC does not match
current configuration in use
Configuration not properly loaded Reinstall current Configuration Profile
304
PLC SYSTEM
FAULT
PLC System Fault
PLC Error
Code
A fault has occurred on the PLC
that prevents the PLC program
from running
Faulty or missing PLC module
PLC is not properly configured
Contact Honeywell Service
305
DATA LOG
FAILED
Unable To Log
Conc Data
Failure to store concentration
data in the database
Local database file has reached size limit
(1GB maximum size for Access
database).
Enable automatic database purge or adjust database purge settings
to purge old data
413
CMD ERR AZ #-
#
Cmd Error Az #-#
Analyzer not responding DAq
Communications
Cable connections Check Ethernet connections
Analyzer missing or powered down Remove analyzer from Configuration Profile