To prevent exposure:
l
Insert ½ Plug (P/N 0235-0168) into the fittings of the pump that was removed for service.
l
Inspect the pump fittings to ensure that the surface is free from score marks and scratches
that could compromise the O-ring seal.
l
Fully seat the removable fitting on the pump after installation.
l
On pumps, verify that the slide plate button extends and mechanically secures the fitting.
With the system running, verify the leak integrity with a small amount of leak test fluid.
6.6.2 Install New Pump
1. Slide pump assembly into enclosure.
2. Push pump completely into enclosure until the release button locks into the bottom of the
enclosure.