8.2.4 Nominal Transport Times
The following table shows the time required for samples to move from the sampling point to the
Vertex M system for various lengths of sample lines.
O.D. 0.25 in. Outside Diameter Tubing
I.D. 0.15 in. (Medium Wall) I.D. 0.190 in. (Thin Wall) I.D.
Length in Feet 100 200 300 400 100 200 300 400
Length in Meters 30 61 91 122 30 61 91 122
# of Analyzers per
line
1
15
sec.
30
sec.
45
sec.
60
sec.
22
sec.
45
sec.
67
sec.
89 sec.
2 8 sec.
15
sec.
11
sec.
22
sec.
34
sec.
45 sec.
3
5.0
sec.
Pressure Limited 7 sec.
15
sec.
Pressure
Limited
Nominal Transport Time in Seconds 1.3-1.8 Liters per Minute per Point (1.5
LPM nominal)
Table 8-2. Nominal Transport Times
Vertex M 168 User Manual