Status indicators include:
l
IDLE - the analyzer is not monitoring or performing maintenance
l
MONT - the analyzer is monitoring
l
LD CC - Load Chemcassette™ routine is in progress
l
PROG - a new application program is being loaded into the analyzer
l
CONF - a new configuration is being loaded into the analyzer
l
COMF - a communications failure between the DAq and the analyzer
l
PYRO-W - a pyrolyzer analyzer is warming up. When pyrolyzer has reached stable
temperature, it will automatically go into monitor
l
FLOW - a flow Auto Balance procedure is in process on the analyzer
The Vertex M system displays only the short names of the location and target gases within each
point block. During normal monitoring, the background color of each block is white. Vertex M
will change the background color of a point as conditions change.
White Normal operation.
Blue Alarms for the point are disabled in the runtime options menu.
Grey Point is disabled in the runtime menu or is disabled due to a fault.
Red Gas concentration exceeded an alarm level.
Purple Disable configuration alarm.
Black Point is not configured for monitoring.
Table 5-1
Alarm Indicators
When the target gas concentration for a point reaches a preset alarm level, Vertex M will display
a W, 1 or 2 in the point block to indicate the severity of the alarm.
Alarm Indicator Concentration Threshold
W (if enabled) Lower Detectable Limit (LDL)
1 Alarm Level 1
2 Alarm Level 2
See "Configure Point" on page60 for information on setting Alarm Level 1 and Alarm Level 2.
See "Other Menu" on page72 for information on alarm default values.
Vertex M 83 User Manual