Chapter 5
Analyze Testing
5-3
Test Sequence
When an individual test is running, the following actions occur:
1. If the test requires the RH (Remote Handset), the DH will first attempt to
communicate with the RH. If it is unable to establish communication, the DH
will display a message that it is looking for the RH and continue trying until the
Analyze test is manually canceled or the RH is found.
2. Once communication with the RH is established, its serial number is checked to
see if current field calibration data exists. If a field calibration has not been
performed in the last 7 days with this RH, an appropriate Calibration
Recommended message will display.
3. If the serial number is valid, the DH will proceed with the specified individual
test.
4. If the test does not require the RH, the DH will run the test and display test
data.
5. Once a test is completed, the test data can be reviewed or printed.
Analyze Test Screens
There are a total of 16 Analyze test screens. They are identical to the
corresponding Autotest screens, with the addition of the Start key located in the
soft key option row. When this key is pressed, the tester begins continuous
execution of the selected test. The Escape key or Cancel key is used to end the
test in progress.
Analyze Test Limitations
Analyze and Autotest test limitations are identical. For a description of test limits
and ranges refer to Appendix C, Measurement Methods and Limits.