REMOTE OPERATION
46882/439 5-13
RF level (source-specific)
RFLV
Set RF Output Level (short form)
:VALUE
Set RF Output Level
Data type :
Decimal Numeric Program Data
Allowed suffices :
Any one of: DBM, DBV, DBMV, DBUV, V, MV, UV OR NV.
Default suffix :
DBM unless changed by UNITS command
:INC
Set RF Level step (dB)
Data type :
Decimal Numeric Program Data
Allowed suffices :
DB only
Default suffix :
DB
:UP
Go UP one step
:DN
Go DOWN one step
:RETN
Return to original setting
:XFER
Transfer current value to be the new setting
:ON
Turn RF Output ON
:OFF
Turn RF Output OFF
Data type :
None
Allowed suffices :
None
Default suffix :
None
:TYPE
Selects EMF or PD for voltage related units
Data type :
Character Program Data (EMF or PD)
Allowed suffices :
None
Default suffix :
None
:UNITS
Select default RF level units.
Data type :
Character Program Data (DBM, DBV, DBMV, DBUV, V, MV or UV)
Allowed suffices :
None
Default suffix :
None
Examples:
RFLV:VALUE 27.3DBM;ON
RFLV:TYPE PD;VALUE 1.23UV
RFLV?
Prepares message containing information on RF Level setting in the following
format:
:RFLV:UNITS <unit>;TYPE <type>;VALUE <nr2>;INC <nr2>;<status>
where: <unit> is character program data defining the default RF level
units (DBM, DBV, DBMV, DBUV, V, MV or UV), <type> is
character program data indicating EMF or PD and <status> is a
program mnemonic indicating whether the RF output is ON or
OFF
Examples:
:RFLV:UNITS DBM;TYPE PD;VALUE 103.5;INC 2.0;ON
:RFLV:UNITS DBV;TYPE EMF;VALUE 83.2;INC 0.5;ON

Table of Contents