REMOTE OPERATION
46882/439 5-19
Amplitude modulation (source-specific)
AM or AM1 or AM2
Set AM Depth (short form)
:DEPTH
Set AM Depth
:INC
Set AM step size
Data type :
Decimal Numeric Program Data
Allowed suffices :
PCT
Default suffix :
PCT
:<src>
Select modulation source where <src> is any one of: INT, EXTAC, EXTALC,
or EXTDC
:ON
Turn AM ON (locally)
:OFF
Turn AM OFF (locally)
:UP
Go UP one step
:DN
Go DOWN one step
:RETN
Return to original setting
:XFER
Transfer current value to be the new setting
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples:
AM:DEPTH 30.5PCT;EXTAC;ON
AM1:DEPTH 40PCT;INT;OFF
:MODF
Set AM modulation oscillator frequency (short form)
:VALUE
Set AM modulation oscillator frequency
:INC
Set AM modulation oscillator frequency step size
Data type :
Decimal Numeric Program Data
Allowed suffices :
Any one of: GHZ, MHZ, KHZ or HZ
Default suffix :
HZ
:UP
Go UP one step
:DN
Go DOWN one step
:RETN
Return to original setting
:XFER
Transfer current value to be the new setting
:SIN
Select sinusoidal waveform
:TRI
Select triangle waveform
:SQR
Select square waveform
Data type :
None
Allowed suffices :
None
Default suffix :
None
:PHASE
Set phase offset of modulation oscillator relative to current phase
Examples:
AM2:MODF:VALUE 15.5KHZ;TRI;INC 500HZ
AM:MODF:PHASE 5DEG