REMOTE OPERATION
46882/439 5-23
Memory protection
MPROT
[not used alone]
:CFRQ
[not used alone]
:START
Set the start of the Carrier Freq Stores memory block which is to be
protected/unprotected.
:STOP
Set the end of the Carrier Freq Stores memory block which is to be
protected/unprotected.
Data type :
Decimal Numeric Program Data
Allowed suffices :
None
Default suffix :
None
:ON
Set memory protection ON (that is, write-protected) for the selected Carrier
Freq Stores memory block.
:OFF
Set memory protection OFF (that is, not write-protected) for the selected
Carrier Freq Stores memory block.
Data type :
None
Allowed suffices :
None
Default suffix :
None
:FULL
[not used alone]
:START
Set the start of the Full Stores memory block which is to be
protected/unprotected.
:STOP
Set the end of the Full Stores memory block which is to be
protected/unprotected.
Data type :
Decimal Numeric Program Data
Allowed suffices :
None
Default suffix :
None
:ON
Set memory protection ON (that is, write-protected) for the selected Full
Stores memory block.
:OFF
Set memory protection OFF (that is, not write-protected) for the selected Full
Stores memory block.
Data type :
None
Allowed suffices :
None
Default suffix :
None
:RAM
[not used alone]
:START
Set the start of the RAM Stores memory block which is to be
protected/unprotected.
:STOP
Set the end of the RAM Stores memory block which is to be
protected/unprotected.
Data type :
Decimal Numeric Program Data
Allowed suffices :
None
Default suffix :
None
:ON
Set memory protection ON (that is, write-protected) for the selected RAM
Stores memory block.
:OFF
Set m
emory protection OFF (that is, not write-protected) for the selected RAM
Stores memory block.
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples:
MPROT:CFRQ:START 50;STOP 99;ON
MPROT:FULL:OFF