EasyManua.ls Logo

IFR 2026

IFR 2026
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UTILITIES
46882/439 3-35
Power-up options (menu 1)
The instrument can power-up in one of two states; with the factory settings or with the settings of
your choice stored in one of the full memory locations. Selection is made as follows:
(1) Press [Power Up Options] to display the Power Up Options Menu shown in Fig. 3-27 below.
This shows the currently selected power-up choice.
B3459
Fig. 3-27 Power up options menu
(2) Press [Factory Power Up] to select the factory power-up settings (these are listed in
Table 3-1 below). Otherwise press [Memory Power Up] to store the current settings as your
selected power-up state.
(3) Press [EXIT] to return to the Utilities Selection Menu 1.
Default settings
The instrument is reset to the factory default settings in the following cases:
(1) At power-up to factory default settings.
(2) Following execution of the
*
RST command.
(3) When [Factory Recall] is pressed on the Memory Recall Menu.
The default settings are shown in Table 3-1.

Table of Contents