EasyManuals Logo

IFR 2026 User Manual

IFR 2026
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
REMOTE OPERATION
5-28 46882/439
BLANK?
Prepares message containing information on the display blanking setting in the
following format:
:BLANK:<state>
where: <state> is program mnemonic indicating whether the blanking is ON
or OFF
Example:
BLANK:OFF
CONTRAST
Sets the LCD contrast, over a scale of 0 to 31.
Data type :
Decimal Numeric Program Data
Allowed suffices :
none
Default suffix :
none
Example:
CONTRAST 16
CONTRAST?
Prepares message containing information on LCD contrast setting in the
following format:
:CONTRAST <nr1>
Example:
:CONTRAST 23
BRIGHTNESS
Sets the LCD brightness to DIM, MEDIUM or BRIGHT.
Data type :
Character Program Data
Allowed suffices :
None
Default suffix :
None
Example:
BRIGHTNESS MEDIUM
BRIGHTNESS?
Prepares message containing information on LCD brightness setting in the
following format:
:BRIGHTNESS <brightness>
where: <brightness> is character program data indicating the brightness
level.
Example:
:BRIGHTNESS BRIGHT
ELAPSED
:RESET
Reset elapsed operating hours to zero
Data type :
None
Allowed suffices :
None
Default suffix :
None
Example:
ELAPSED:RESET
ELAPSED?
Prepares message containing information on elapsed operating hours since last
reset. Fractional part is in 15 minute intervals (0.25, 0.50, 0.75). Format is as
follows:
<nr2>
Example:
454.50
OPER?
Prepares message containing information on total operating hours. Fractional
part is in 15 minute intervals (0.25, 0.50, 0.75). Format is as follows::
<nr2>
Example:
1453.00
POWUP
[not used alone]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IFR 2026 and is the answer not in the manual?

IFR 2026 Specifications

General IconGeneral
BrandIFR
Model2026
CategoryInverter
LanguageEnglish