7 | Warning and error messages INFICON
104 / 152 UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
Type Notification Possible sources of
error
Remedy
W360 Preamplifier output signal
too low
Ion source defective Contact customer service
Mass spectrometer
contaminated
Contact customer service
W361 The offset voltage of the
preamplifier is too high
Preamplifier defective Contact customer service
W362 Preamplifier range error Preamplifier defective Contact customer service
MSB defective Contact customer service
W374 Background in stand-by out
of range
Mass spectrometer
contaminated
Check the tightness of the connection
between the leak detector and the test
object
Reduce contamination in the environment
where the device is located
Use the 'Regeneration' function in the
'Vacuum' operation mode
Ion source or
preamplifier defective
Contact customer service
W390 500 GOhm amplifier resistor
out of range
Preamplifier defective Contact customer service
MSB defective Contact customer service
E400 TMP error message Unhandled error
message from the TMP
Contact customer service
W401 TMP warning message Unhandled warning
message from the TMP
Contact customer service
E402 No communication with
turbo pump converter
TMP cable defective or
not connected
Contact customer service
TMP converter
defective
Contact customer service
MSB defective Contact customer service
E404 TMP current consumption
too high
Pressure in the TMP is
too high
Contact customer service
W405 No TMP run-up Inlet pressure during
run-up too high
Check that the backing pump is working
during run-up
Check that the pressure drops during run-
up
Contact customer service
TMP bearing damage Contact customer service