INFICON Warning and error messages | 7
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 99 / 152
Type Notification Possible sources of
error
Remedy
W153 Operating unit software
version is obsolete
A more up-to-date
operating unit software
is available. For
trouble-free operation,
it is recommended to
update the operating
unit software.
• Contact the customer service for the
latest operating unit software
E167 Run-up blocked, the
'Interlock' function is active
The dongle is not
connected to the
ACCESSORIES port
• Plug the dongle into the ACCESSORIES
port
The dongle connected
to the ACCESSORIES
port is defective
• Use another module if possible
The 'Interlock' function
was activated
unintentionally
• Deactivate the 'Interlock' function
W171 CU1000 not supported A CU1000 cannot be
used with this device
• Disconnect the CU1000 from this device
W201 Voltage U24V_MSB too low Malfunction of 24V
power supply unit
• Contact customer service
Short circuit or
overload in the 24V
supply
• Contact customer service
W202 Voltage U24V_MSB too high Malfunction of 24V
power supply unit
• Contact customer service
W203 24V valve supply voltage out
of range
Malfunction of a valve • Contact customer service
Short circuit or
overload in the 24V
valve supply
• Contact customer service
W206 24V operating unit supply
voltage out of range
Malfunction of
operating unit
• Contact customer service
Short circuit or
overload in the 24V
operating unit supply
• Contact customer service
W207 12V internal supply voltage
out of range
Short circuit or
overload in the 12V
supply
• Contact customer service