7 | Warning and error messages INFICON
100 / 152 UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
Type Notification Possible sources of
error
Remedy
W208 24V fan supply voltage out
of range
Malfunction of a fan Contact customer service
Short circuit or
overload in the 24V fan
supply
Contact customer service
W209 24V DIV1 supply voltage out
of range
Short circuit or
overload in the 24V
DIV1 supply
Contact customer service
W210 24V DIV2 supply voltage out
of range
Short circuit or
overload in the 24V
DIV2 supply
Contact customer service
W211 5V internal supply voltage
out of range
Short circuit or
overload in the internal
5V supply
Contact customer service
W212 5V sniffer line supply voltage
out of range
Sniffer line defective Use another sniffer line, if possible
Contact customer service
Short circuit or
overload in the 5V
sniffer line supply
Contact customer service
W221 Internal voltage 24V_RC
voltage out of range
The remote control is
defective
Use another remote control, if possible
The cable connected to
REMOTE CONTROL
connector is defective
Use another cable, if possible
The accessory
connected to
ACCESSORIES
connector is defective
Use another accessory, if possible
The cable connected to
ACCESSORIES
connector is defective
Use another cable, if possible
Short circuit or
overload in the
24V_RC supply
Contact customer service