INFICON Warning and error messages | 7
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 101 / 152
Type Notification Possible sources of
error
Remedy
W222 Internal voltage 24V_IO
voltage out of range
The module connected
to LD connector is
defective
Use another module, if possible
The cable connected to
LD connector is
defective
Use another cable, if possible
Short circuit or
overload in the 24V_IO
supply
Contact customer service
W223 Internal voltage 24V_TMP1
voltage out of range
Short circuit or
overload in the
24V_TMP1 supply
Contact customer service
W224 Internal voltage
24V_PRESSURE (Pirani)
voltage out of range
Short circuit or
overload in the
24V_PRESSURE
(Pirani) supply
Contact customer service
W240 Voltage +15V out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
W241 Voltage -15V out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
E242 +15V or -15V voltage
shorted
Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
W250 REF5V voltage out of range Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service
E252 REF5V voltage shorted Short circuit at
preamplifier
Contact customer service
VI-Board or MSB
defective
Contact customer service