2.2 Electrical data
Calculator data
Typical accuracy
Calculator: E
c
±
(0,15 + 2/
∆Θ
) % Sensor set: E
t
±
(0,4 + 4/
∆Θ
) %
Display
LCD 7 or 8 digits, digit height 8,2 mm
Resolutions
9999,999 99999,99 999999,9 9999999
99999,999 999999,99 9999999,9 99999999
Energy units
MWh – kWh GJ
Data logger (EEPROM),
Programmable
Logging intervals: From one minute to one year
Logger content: All registers can be selected
Standard logger profile: 20 years, 36 months, 460 days, 72 hours
Info logger (EEPROM)
50 info codes
Clock/calendar
(with back-up battery)
Clock, calendar, leap year compensation, target date
Clock deviation
Unadjusted by AMR: Less than 15 min./year from legal time
Adjusted by AMR every 48 hrs: Less than 7 sec. from legal time
Summer/winter time (DST)
Programmable under country code.
The function can be disabled so that ”technical normal time” is used
Data communication
KMP protocol with CRC16 used for optical communication
as well as modules.
Power of temperature
sensors
<
10
µ
W RMS
Supply voltage
3,6 VDC ± 0,1 VDC
Battery
3,65 VDC, D-cell lithium
3,65 VDC, 2 x A cells lithium
3,65 VDC, C-cell lithium
1
Replacement interval
- Wall mounted
- Mounted on flow sensor
16 years @ t
BAT
< 30 °C
14 years @ t
BAT
< 40 °C
9 years @ t
BAT
< 30 °C
8 years @ t
BAT
< 40 °C
See paragraph 10.4 for further information.
Back-up battery
(for real time clock)
3,0 VDC, BR-cell lithium
Lifetime: 20 years @ t
BAT
< 40 °C with connected battery/supply or
4 years @ t
BAT
< 40 °C without connected battery/supply
Mains supply
230 VAC +15/-30 %, 50/60 Hz
24 VAC ±50 %, 50/60 Hz
Insulation voltage
3,75 kV
Power consumption
< 1 W
Back-up supply
Integral SuperCap eliminates interruptions due to short-term power failures
(Supply modules type 403-xxxxxxxxxxx7 and -8 only)
EMC data
Fulfils EN 1434 class A and C (MID class E1 and E2)
1
For especially current-demanding applications, such as battery operation of an NB-IoT module. This battery package
includes a lithium C-cell as well as a special capacitor that helps the battery to deliver high pulse currents.

Table of Contents

Other manuals for Kamstrup MULTICAL 403

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kamstrup MULTICAL 403 and is the answer not in the manual?

Kamstrup MULTICAL 403 Specifications

General IconGeneral
TypeUltrasonic Heat Meter
Flow SensorUltrasonic
Pressure RatingPN16
Battery LifeUp to 16 years
ApprovalMID Class 2
DisplayLCD
CommunicationM-Bus, Wireless M-Bus, Modbus, BACnet
Power SupplyBattery
AccuracyClass 2
Protection ClassIP54
Housing MaterialComposite
Flow Rate Range0.6 m3/h to 15 m3/h