Kavo 3D eXam ® Operators’ Manual
K990400 16 April 2008
13-2
Sensor Front Panel Attenuation Value: Less than 1mm of aluminum
equivalent
Gray Scale: 14 bit
Voxel Size: 0.4/0.3/0.25/0.2 mm
Image Acquisition: Single 360 degree rotation (maximum)
Scan Time: 26.9/8.9 seconds
Field of View: (standard) 6.50 inches diameter x 5.33 inches height
(maximum) (16.5 cm diameter x 13.5 cm height (maximum)
Extended Field of View: (optional) 9.15 inches diameter x 6.67 inches
height maximum) (23.2 cm diameter x 17 cm height (maximum))
Note: Maximum values can be collimated down.
Primary Reconstruction: Less than 2 minutes for 26.9 second scan @
0.4 voxel
Secondary Reconstruction: Real Time
Power Requirements
The Scanner requires a Dedicated Line. A Surge Protector is recommended.
Line Voltage: 115VAC, 200VAC or 230VAC (Factory Set)
Line Voltage Regulation requirement: + 10%
Line Current: 10 Amps (115V), 7.5 Amps (200V) or 5 Amps (230V)
Line Frequency: 50 Hz / 60 Hz
Phase: Single
Main Circuit Breaker: 10 Amps (115V), 7.5 Amps (200V), or 5 Amps
(230V)
Power Cords
Power Cords are not supplied with the system. Three Power Cords are
required, one each for the Gantry, Computer, and Monitor.
Power Cord Specifications:
10 Amp, 220-240VAC, 50 Hz / 60 Hz, 16/3 Gauge (minimum) Length
6 to 12 ft. (2m to 4m), IEC-320-C13 Connector End.