K990400 16 April 2008
Product Information
13-9
Cleaning
Cleaning the equipment frequently, especially if corroding
chemicals are present is a function of the Operator. Unless otherwise
instructed, use a cloth moistened with warm water and mild soap. Do
not use strong cleaners and solvents as these may damage the finish.
Be careful when cleaning to avoid liquid leaking inside the Gantry.
USE EXTREME CARE WHEN WORKING INSIDE THE
SYSTEM. VOLTAGES ARE PRESENT WHEN THE SYSTEM
COVERS ARE OFF. ACCIDENTAL CONTACT WITH THESE
VOLTAGES MAY CAUSE SERIOUS INJURY OR DEATH.
Planned Maintenance Checklist
Perform Calibrations
Panel Calibration
Collimator Calibration
Geometry Calibration
Check Detector Pivot (displayed under Geometry, both Portrait and Landscape)
Ensure Detector Pivot is between -0.10 & 0.10. If measurements are not within
tolerance, call Service to have Receptor Panel leveled.
Peform QA Water Phantom Tests
Noise Level Test (Landscape Mode)
Uniformity Test (Landscape Mode)
Noise Level Test (Portrait Mode)
Uniformity Test (Portrait Mode)
PAN Phantom Test
Radiation Ouput Test (performed by qualitfied Personnel, per local requirements)
Check Chair Alignments
Check Chair Center Alignment (using Chair Center Locator)
Check Chair Level (using Chair Center Locator)