Model 2750 Multimeter/Switch System User’s Manual Rel, Math, Ratio, Channel Average, dB 5-5
Commands
1
Description Default
Rel commands for ACI
[SENSe[1]] Optional root command.
:CURRent:AC:REFerence <n> [, <clist>] Specify rel value; <n> = -3.1 to 3.1 (A). 0
:CURRent:AC:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF. OFF
:CURRent:AC:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for
Ω2
[SENSe[1]] Optional root command.
:RESistance:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 120e6 (Ω). 0
:RESistance:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:RESistance:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for
Ω4
[SENSe[1]] Optional root command.
:FRESistance:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 120e6 (Ω). 0
:FRESistance:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:FRESistance:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for TEMP
2
[SENSe[1]] Optional root command.
:TEMPerature:REFerence <n> [, <clist>] Specify rel value; <n> = -200 to 1820 (°C). 0
:TEMPerature:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:TEMPerature:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for FREQ
[SENSe[1]] Optional root command.
:FREQuency:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 1.5e7 (Hz). 0
:FREQuency:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:FREQuency:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Table 5-1 (continued)
Rel commands
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM