15-6 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
Table 15-2
DISPlay command summary
Command Description
Default
parameter Ref SCPI
DISPlay (see Note) Sec 1
[:WINDow[1]]
:TEXT Path to control user text messages.
:DATA <a> Define ASCII message “a”
(up to 12 characters).
(none)
:DATA? Query text message.
:STATe <b> Enable or disable message mode. OFF
:STATe? Query text message state.
:ENABle <b> Enable or disable the front panel display. OFF
:ENABle? Query state of the display.
Note: *RST and SYSTem:PRESet have no effect on commands in this subsystem. The listed defaults are power-on defaults.
Table 15-3
FORMat command summary
Command Description
Default
parameter Ref SCPI
FORMat Sec 14
[:DATA] <type>
[,<length>]
Select data format (ASCii, SREal, or DREal). ASCii
[:DATA]? Query data format.
:ELEMents <item list> Specify data elements (READing, CHANnel,
UNITs, RNUMber, TSTamp, and LIMits).
(see Note)
:ELEMents? Query data elements.
:BORDer <name> Select binary byte order (NORMal or
SWAPped).
SWAPped
:BORDer? Query byte order.
Note: The SYSTem:PRESet and *RST default is READ, UNIT, RNUM, and TST.