EasyManuals Logo

Keithley 2750 User Manual

Keithley 2750
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #353 background imageLoading...
Page #353 background image
15-6 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
Table 15-2
DISPlay command summary
Command Description
Default
parameter Ref SCPI
DISPlay (see Note) Sec 1
[:WINDow[1]]
:TEXT Path to control user text messages.
:DATA <a> Define ASCII message “a”
(up to 12 characters).
(none)
:DATA? Query text message.
:STATe <b> Enable or disable message mode. OFF
:STATe? Query text message state.
:ENABle <b> Enable or disable the front panel display. OFF
:ENABle? Query state of the display.
Note: *RST and SYSTem:PRESet have no effect on commands in this subsystem. The listed defaults are power-on defaults.
Table 15-3
FORMat command summary
Command Description
Default
parameter Ref SCPI
FORMat Sec 14
[:DATA] <type>
[,<length>]
Select data format (ASCii, SREal, or DREal). ASCii
[:DATA]? Query data format.
:ELEMents <item list> Specify data elements (READing, CHANnel,
UNITs, RNUMber, TSTamp, and LIMits).
(see Note)
:ELEMents? Query data elements.
:BORDer <name> Select binary byte order (NORMal or
SWAPped).
SWAPped
:BORDer? Query byte order.
Note: The SYSTem:PRESet and *RST default is READ, UNIT, RNUM, and TST.
2750-900-01.book Page 6 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals