EasyManuals Logo

Keithley 2750 User Manual

Keithley 2750
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #361 background imageLoading...
Page #361 background image
15-14 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
[SENSe[1]]
:FRESistance Path to configure four-wire resistance. Sec 3
:APERture <NRf>
[, <clist>]
Set integration rate in seconds (60Hz,
1.67e-4 to 1; 50Hz, 2e-4 to 1).
(Note 2) Sec 4
:APERture? [<clist>] Query aperture integration rate.
:NPLCycles <n>[, <clist>] Set integration rate in line cycles (60Hz,
0.01 to 60; 50Hz, 0.01 to 50).
5.0 Sec 4
:NPLCycles? [<clist>] Query line cycle integration rate.
:RANGe Path to set measurement range: Sec 4
[:UPPer] <n> [, <clist>] Select range (0 to 120e6). 120e6
[:UPPer]? [<clist>] Query range.
:AUTO <b> [, <clist>] Enable or disable auto range. ON
:AUTO? [<clist>] Query state of auto range.
:DIGits <n> [, <clist>] Specify measurement resolution (4 to 7). 7 Sec 4
:DIGits? [<clist>] Query resolution.
:REFerence <n> [, <clist>] Specify reference (0 to 120e6). 0 Sec 5
:STATe <b> [, <clist>] Enable or disable reference. OFF
:STATe? [<clist>] Query state of reference.
:ACQuire [, <clist>] Use input signal as reference.
:REFerence? [<clist>] Query reference value.
:AVERage Path to configure and control filter: Sec 4
:TCONtrol <name> Select filter type: (MOVing or REPeat). (Note 3)
:TCONtrol? Query filter type.
:WINDow <NRf> Set filter window in % of range
(0 to 10).
0.1
:WINDow? Query filter window.
:COUNt <n> [, <clist>] Specify filter count (1 to 100). 10
:COUNt? [<clist>] Query filter count.
[:STATe] <b> [, <clist>] Enable or disable filter. (Note 4)
[:STATe]? [<clist>] Query state of digital filter.
:OCOMpensated <b>
[, <clist>]
Enable or disable offset compensation. OFF Sec 3
:OCOMpensated? [<clist>] Query state of offset compensation.
:DCIRcuit <b> [, <clist>] Enable or disable dry circuit ohms. OFF Sec 3
:DCIRcuit? [, <clist>] Query state of dry circuit ohms.
Table 15-5 (continued)
SENSe command summary
Command Description
Default
parameter Ref SCPI
2750-900-01.book Page 14 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A