15-26 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
Table 15-10
UNIT command summary
Command Description
Default
parameter Ref SCPI
UNIT
:TEMPerature <name> Select temperature units (C, CEL, F,
FAR, or K).
C Sec 3
:TEMPerature? Query temperature units.
:VOLTage Path to configure voltage units. Sec 5
[:DC] <name>
[, <clist>]
Select DCV measurement units (V or DB). V
:DB Path to set DB reference voltage:
:REFerence <n> Specify reference in volts (1e-7 to 1000). 1
:REFerence? Query reference.
[:DC]? [<clist>] Query DCV units.
:AC <name> [, <clist>] Select ACV measurement units (V or DB). V
:DB Path to set DB reference voltage.
:REFerence <n> Specify reference in volts (le-7 to 1000). 1
:REFerence? Query DB reference.
:AC? [<clist>] Query ACV units.