Model 2750 Multimeter/Switch System User’s Manual SCPI Reference Tables 15-25
Table 15-9
Trigger command summary
Command Description
Default
parameter Ref SCPI
INITiate Subsystem command path: Sec 8
[:IMMediate] Initiate one trigger cycle.
:CONTinuous <b> Enable or disable continuous initiation. (Note 1)
:CONTinuous? Query continuous initiation.
ABORt Reset trigger system.
TRIGger[:SEQuence[1]] Path to program Trigger Layer:
:COUNt <n> Set measure count (1 to 110000, or INFinity). (Note 2)
:COUNt? Query measure count.
:DELay <n> Set delay (0 to 999999.999 sec). 0
:AUTO <b> Enable or disable auto delay. OFF
:AUTO? Query state of delay.
:DELay? Query delay.
:SOURce <name> Select control source (IMMediate, TIMer,
MANual, BUS, or EXTernal).
IMM
:SOURce? Query control source.
:TIMer <n> Set timer interval (0.001 to 999999.999 sec). 0.1
:TIMer? Request the programmed timer interval.
:SIGNal Loop around control source.
SAMPle
:COUNt <NRf> Specify sample count (1 to 110000). 1
:COUNt? Query sample count.
Notes:
1. Defaults for continuous initiation:
SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. Defaults for count:
SYSTem:PRESet sets the count to INF (infinity).
*RST sets the count to 1.
2750-900-01.book Page 25 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
BrandKeithley
Model2750
CategorySwitch
LanguageEnglish

Related product manuals