EasyManua.ls Logo

Keithley 2750 - Page 371

Keithley 2750
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15-24 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
Table 15-8
TRACe command summary
Command Description
Default
parameter* Ref SCPI
TRACe|:DATA Use TRACe or DATA as root command. Sec 6
:CLEar Path to clear the buffer.
[:IMMediate] Clear the buffer.
:AUTO <b> Enable or disable buffer auto-clear. ON
:AUTO? Query state of buffer auto-clear.
:FREE? Query bytes available and bytes in use.
:POINts <NRf> Specify size of buffer (2 to 110000). 100
:POINts? Query buffer size.
:NOTify <NRf> Specify number of stored readings that will set
Trace Notify bit (B6) of measurement event
register (2 to 109999). Must be less than
TRACe:POINts value.
50
:NOTify? Query trace notify value.
:NEXT? Query buffer location for next stored reading.
:TSTamp Path to set timestamp format:
:FORMat <name> Select timestamp format (ABSolute or DELTa). ABS
:FORMat? Query timestamp format.
TYPE? Query timestamp type for readings presently in
buffer.
:FEED <name> Select source of readings (SENSe[1],
CALCulate[1], or NONE).
CALC
:CONTrol <name> Select buffer control mode (NEVer, NEXT, or
ALWays).
NEV
:CONTrol? Query buffer control mode.
:FEED? Query source of readings for buffer.
:DATA? Read all readings in the buffer.
:DATA:SELected?
<start>, <count>
Specify readings to be returned; specify starting
point (first reading is #0) and number of readings
(count).
*SYSTem:PRESet and *RST have no effect on commands in this subsystem. The listed defaults are defaults set at the factory.
2750-900-01.book Page 24 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Related product manuals