15-24 SCPI Reference Tables Model 2750 Multimeter/Switch System User’s Manual
Table 15-8
TRACe command summary
Command Description
Default
parameter* Ref SCPI
TRACe|:DATA Use TRACe or DATA as root command. Sec 6
:CLEar Path to clear the buffer.
[:IMMediate] Clear the buffer.
:AUTO <b> Enable or disable buffer auto-clear. ON
:AUTO? Query state of buffer auto-clear.
:FREE? Query bytes available and bytes in use.
:POINts <NRf> Specify size of buffer (2 to 110000). 100
:POINts? Query buffer size.
:NOTify <NRf> Specify number of stored readings that will set
Trace Notify bit (B6) of measurement event
register (2 to 109999). Must be less than
TRACe:POINts value.
50
:NOTify? Query trace notify value.
:NEXT? Query buffer location for next stored reading.
:TSTamp Path to set timestamp format:
:FORMat <name> Select timestamp format (ABSolute or DELTa). ABS
:FORMat? Query timestamp format.
TYPE? Query timestamp type for readings presently in
buffer.
:FEED <name> Select source of readings (SENSe[1],
CALCulate[1], or NONE).
CALC
:CONTrol <name> Select buffer control mode (NEVer, NEXT, or
ALWays).
NEV
:CONTrol? Query buffer control mode.
:FEED? Query source of readings for buffer.
:DATA? Read all readings in the buffer.
:DATA:SELected?
<start>, <count>
Specify readings to be returned; specify starting
point (first reading is #0) and number of readings
(count).
*SYSTem:PRESet and *RST have no effect on commands in this subsystem. The listed defaults are defaults set at the factory.
2750-900-01.book Page 24 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals