EasyManuals Logo

Keithley 2750 User Manual

Keithley 2750
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #421 background imageLoading...
Page #421 background image
F-4 Temperature Equations Model 2750 Multimeter/Switch System User’s Manual
Table F-4
Type K inverse function polynomial
-200°C to 0°C
(-5,891µV to 0µV)
0°C to 500°C
(0µV to 20,644µV)
500°C to 1,372°C
(20,644µV to 54,886µV)
c
0
= 0.0 0.0 -1.318 058 × 10
2
c
1
= 2.517 346 2 × 10
-2
2.508 355 2 × 10
-2
4.830 222 × 10
-2
c
2
= -1.166 287 8 × 10
-6
7.860 106 2 × 10
-8
-1.646 031 × 10
-6
c
3
= -1.083 363 8 × 10
-9
-2.503 131 2 × 10
-10
5.464 731 × 10
-11
c
4
= -8.977 354 0 × 10
-13
8.315 270 2 × 10
-14
-9.650 715 × 10
-16
c
5
= -3.734 237 7 × 10
-16
-1.228 034 2 × 10
-17
8.802 193 × 10
-21
c
6
= -8.663 264 3 × 10
-20
9.804 036 2 × 10
-22
-3.110 810 × 10
-26
c
7
= -1.045 059 8 × 10
-23
-4.413 030 2 × 10
-26
c
8
= -5.192 057 7 × 10
-28
1.057 734 2 × 10
-30
c
9
= -1.052 755 2 × 10
-35
Error: 0.04°C to -0.02°C 0.04°C to -0.05°C 0.06°C to -0.05°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
Table F-5
Type N inverse function polynomial
-200°C to 0°C
(-3,990µV to 0µV)
0°C to 600°C
(0uV to 20,613µV)
600°C to 1,300°C
(20,613µV to 47,513µV)
c
0
= 0.0 0.0 1.972 485 × 10
1
c
1
= 3.843 684 7 × 10
-2
3.868 96 × 10
-2
3.300 943 × 10
-2
c
2
= 1.101 048 5 × 10
-6
-1.082 67 × 10
-6
-3.915 159 × 10
-7
c
3
= 5.222 931 2 × 10
-9
4.702 05 × 10
-11
9.855 391 × 10
-12
c
4
= 7.206 052 5 × 10
-12
-2.121 69 × 10
-18
-1.274 371 × 10
-16
c
5
= 5.848 858 6 × 10
-15
-1.172 72 × 10
-19
7.767 022 × 10
-22
c
6
= 2.775 491 6 × 10
-18
5.392 80 × 10
-24
c
7
= 7.707 516 6 × 10
-22
-7.981 56 × 10
-29
c
8
= 1.158 266 5 × 10
-25
c
9
= 7.313 886 8 × 10
-30
Error: 0.03°C to -0.02°C 0.03°C to -0.02°C 0.02°C to -0.04°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
2750-900-01.book Page 4 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A