EasyManuals Logo

Keithley 2750 User Manual

Keithley 2750
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #422 background imageLoading...
Page #422 background image
Model 2750 Multimeter/Switch System User’s Manual Temperature Equations F-5
Table F-6
Type R inverse function polynomial
-50°C to 250°C
(-226µV to
1,923µV)
250°C to 1,200°C
(1,923µV to
13,228µV)
1,064°C to 1,664.5°C
(11,361µV to
19,739µV)
1,664.5°C to 1,768.1°C
(19,739µV to
21,103µV)
c
0
= 0.0 1.334 584 505 × 10
1
-8.199 599 416 × 10
1
3.406 177 836 × 10
4
c
1
= 1.889 138 0 × 10
-1
1.472 644 573 × 10
-1
1.553 962 042 × 10
-1
-7.023 729 171
c
2
= -9.383 529 0 × 10
-5
-1.844 024 844 × 10
-5
-8.342 197 663 × 10
-6
5.582 903 813 × 10
-4
c
3
= 1.306 861 9 × 10
-7
4.031 129 726 × 10
-9
4.279 433 549 × 10
-10
-1.952 394 635 × 10
-8
c
4
= -2.270 358 0 × 10
-10
-6.249 428 360 × 10
-13
-1.191 577 910 × 10
-14
2.560 740 231 × 10
-13
c
5
= 3.514 565 9 × 10
-13
6.468 412 046 × 10
-17
1.492 290 091 × 10
-19
c
6
= -3.895 390 0 × 10
-16
-4.458 750 426 × 10
-21
c
7
= 2.823 947 1 × 10
-19
1.994 710 149 × 10
-25
c
8
= -1.260 728 1 × 10
-22
-5.313 401 790 × 10
-30
c
9
= 3.135 361 1 × 10
-26
6.481 976 217 × 10
-35
c
10
= -3.318 776 9 × 10
-30
Error: 0.02°C to -0.02°C 0.005°C to -0.005°C 0.001°C to -0.0005°C 0.002°C to -0.001°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
Table F-7
Type S inverse function polynomial
-50°C to 250°C
(-235µV to 1,874µV)
250°C to 1,200°C
(1,874µV to 11,950µV)
1,064°C to 1,664.5°C
(10,332µV to 17,536µV)
1,664.5°C to 1,768.1°C
(17,536µV to 18,693µV)
c
0
= 0.0 1.291 509 199 × 10
1
-8.087 801 117 × 10
1
5.333 875 126 × 10
4
c
1
= 1.849 494 60 × 10
-1
1.466 298 863 × 10
-1
1.621 573 104 × 10
-1
-1.235 892 298 × 10
1
c
2
= -8.005 050 62 × 10
-5
-1.534 713 402 × 10
-5
-8.536 869 453 × 10
-6
1.092 657 613 × 10
-3
c
3
= 1.022 374 30 × 10
-7
3.145 945 973 × 10
-9
4.719 686 976 × 10
-10
-4.265 693 686 × 10
-8
c
4
= -1.522 485 92 × 10
-10
-4.163 257 839 × 10
-13
-1.441 693 666 × 10
-14
6.247 205 420 × 10
-13
c
5
= 1.888 213 43 × 10
-13
3.187 963 771 × 10
-17
2.081 618 890 × 10
-19
c
6
= -1.590 859 41 × 10
-16
-1.291 637 500 × 10
-21
c
7
= 8.230 278 80 × 10
-20
2.183 475 087 × 10
-26
c
8
= -2.341 819 44 × 10
-23
-1.447 379 511 × 10
-31
c
9
= 2.797 862 60 × 10
-27
8.211 272 125 × 10
-36
Error: 0.02°C to -0.02°C 0.01°C to -0.01°C 0.0002°C to -0.0002°C 0.002°C to -0.002°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C.
E is the measured voltage in microvolts.
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals