7) Confirm the tilt error and cornerload error levels according to "4.2 Adjusting the Tilt Error" and "4.3
Adjusting the Cornerload Error." If the error exceeds allowable levels, adjust the tilt or cornerload.
Enter the attached EEPROM (temperature coefficient) data into the replacement unit assembly (see
4 *2).
Temperature Coefficients at Address Numbers 00CA to 0131 00
CA to 0011, 05A0 to 05AF,
an
d 0112 to 0131
8)
22 DBS-SH-e-1210