9
.
5
He
a
te
r
D
o
es N
o
t
C
o
m
e
O
N
He
a
te
r
does
n
o
t switch ON.
No
Is the cumulative
time less than 5000
hours?
Y
es
Is the resistance
greater
than 100kΩ ?
Y
es
No
Y
es
Is the jun
c
tion
cable
broken?
No
Y
es
Is the the
r
m
a
l
fuse
blown?
No
Replace
the
power board
assembly.
Replace
the
main board
assembly.
Replace
the
thermal
fuse.
Check
the
thermal
fuse.
Replace
the
junction
cable.
Check
the
junction
cable (H7).
Replace
the heater.
Check
the
heater
(S5).
Replace
the heater.
Open
the
service menu and
check
the heater time
count.
64 DBS-SH-e-1210