4.
H
rdw
e
Ad
ustment
Replacing the following components may alter balance performance. After replacement, check
he
e
mance and ma
e ad
us
men
si
necessa
.
*1) When the main board assembly (B2) has been replaced, mount the EEPROM (M4 and M5)that was
included with the original main board assembly (B2).
*2) Input the included EEPROM data into the balance (temperature coefficients at address numbers
00CA - 0131, and 05A0 - 05AF).
*3) See 6.2 Data Edit (EDIT) Operations and set address numbers 0234 - 0235 and 03DE - 03E7 to "0."
(Error Log Reset)
Replacing the following components may alter heater performance. After replacement, make adjustments if
necessary.
*4) See 6.2 Data Edit (EDIT) Operations and set add
ess numbe
s02
4-02
5and 0
DE - 0
E7 to "0."
*5) Replace the thermistor (H10) also, when replacing the thermal fuse. (H6)
DBS-SH-e-1210 41