8.2
E
r
r
o
r
D
i
s
p
l
a
y
Failed
to
read from
the
EEPROM.
Measurement was
attempted without pressing the
[TARE] key.
(2)
Temperature exceeded the setting value
and increased
rapidly
(at least 15
°C per 10
value
exceeded 2 %
of the
mass
of the
standard calibration
weight.
Error
No.
Name
Description
(Warning) OL Overload
Weighing capacity
is
exceeded.
(Warning) -OL Underload
Pan supporter is
missing.
ERR.
OL
Overload
During
Measurement
Mass
exceeded
weighing
capacity
during measurement.
ERR.
-OL
Underload
During
Measurement
Pan supporter was
removed during measurement.
ERR.
001
Mass
Temperature Sensor
(TAD)
Open
Circuit
Circuit broke
in
the mass
temperature sensor.
ERR.
002
Mass
Temperature Sensor
(TAD)
Short
Circuit
Short circuit occurred
in
the mass
temperature sensor.
ERR.
005
EEPROM Data
Error
Invalid
value was
written
to the EEPROM.
ERR.
100
LidOpen1
Minute Error
Lid was
open for
1
minute
or more during a measurement (when heated).
ERR.
101 Thermistor Open
Circuit broke in the
thermistor.
ERR.
102 Thermistor Short
Shortcircuitoccurredinthe
thermistor.
ERR.
110
Lid Switch
Error
Mass
increased even
though
the lid is
not open.
ERR.
111
TARE
Error
Measurement was not started
within 3
minutes
of
pressing the [TARE]
key.
ERR.
121
Heater Failure
Temperature
did not
rise during
measurements.
Starting temperature differential (difference between temperature
when
heating starts
and
temperature setting) was
at least 60
degrees
and
temperature
did not
increase by
15
degrees within
30
seconds
(5
degrees
within 60
seconds
in the slow
mode,
however).
Decision
is
aborted
if
the
lid is
open before decision.
Also,
decision is
aborted if
differential is
less
than 60
degrees.
ERR.
122
Heater
Error
Not during measurement (when heater
is
OFF)
(1)
Sensor
detected
temperature rose
(at least 20
°
Cper
10 sec).
During measurement
sec).
(3)
Temperature increased
after heating
ended (at least 16
°
C
per 10 sec).
ERR.
123
Heating
Error
Thermistor temperature exceeded 270
°
C
ERR.
124 M Error M
did
not reach
the setting
value
foratleast4hours.
ERR.
200
Power
Supply
Error
Power
supply
voltage
range
setting in
CPU is
inconsistent
with the
actual power
supply.
ERR.
201
Unknown
Frequency Error
Power
supply frequency could
not be determined.
ERR.
202
Unknown
Voltage
Power
supply
voltage could
not be determined.
COM.
ERR
Communications
Error
Unrecognized command
was
received during command reception.
ERR.
C01
Instability
Error
Unstable state continued for at least 1 minute during mass
calibration.
ERR.
C04
Span
Error
Difference between the measured span value and specified standard calibration
weight
ABORT Abort
Moisture analyzer
function
was
aborted.
58 DBS-SH-e-1210