Y
es
Is the Pt-Ni band
(U17) broken or
twisted?
No
Is the force coil
(L1) contacting the
magnet assembly
(U9)?
Y
es
No
Y
es
Is the
f
o
r
c
e
c
oil
(L1) leaking?
No
Is the detector
assembly (U11)
and its adjustment
normal?
No
Y
es
No
Is the
unit
assembly
(U)
normal?
Y
es
Replace
the
unit assembly
(U).
Replace
the
analog board assembly
(B3).
Replace
the
EEPROM
on
the main
board
assembly
(B2).
Replace with
a
properly functioning
unit
assembly
(U).
Readjust the detector assembly (U11)
height.
Replace
the
detector assembly
(U11).
Reattach the
unit
assembly
(U).
Replace
the force coil
assembly
(L1).
Readjust
the
force coil
(L1)
attachment
position.
Remove
the
unit assembly (U) and then magnetic
lids.
Reinstall
the
Pt-Ni band
(U17).