Page #249 background imageLoading...
Page #249 background image
Utility Settings 20
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 249
To establish a LAN connection
Automatic
Configuration
1 Press [Utility] > I/O.
2 Press the LAN Settings softkey.
3 Press the Config softkey; then, turn the Entry knob to select Automatic, and press
the softkey again to enable it.
If your network supports DHCP or AutoIP, enabling Automatic lets the
oscilloscope use those services to get its LAN configuration settings
4 If your network provides Dynamic DNS, you can enable the Dynamic DNS option
to let the oscilloscope register its hostname and use the DNS server for name
resolution.
5 You can enable the Multicast DNS option to let the oscilloscope use Multicast
DNS for name resolution on small networks without a conventional DNS server.
6 Connect the oscilloscope to the local area network (LAN) by inserting the LAN
cable into the "LAN" port on the rear panel of the oscilloscope.
In a few moments the oscilloscope will connect to the network automatically.
If the oscilloscope does not automatically connect to the network, press [Utility]
> I/O > LAN Reset. In a few moments the oscilloscope will connect to the
network.
Manual
Configuration
1 Get the oscilloscope's network parameters (host name, IP address, subnet
mask, gateway IP, DNS IP, etc.) from your network administrator.
2 Press [Utility] > I/O.
3 Press the LAN Settings softkey.
4 Press the Config softkey; then, turn the Entry knob to select Automatic, and press
the softkey again to disable it.
If Automatic is not enabled, the oscilloscope's LAN configuration must be set
up manually using the Addresses and Host name softkeys
NOTE
Any time you modify the oscilloscope's hostname, it breaks the connection between the
oscilloscope and the LAN. You need to re-establish communication to the oscilloscope using
the new hostname.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies MSO-X 2002A and is the answer not in the manual?

Keysight Technologies MSO-X 2002A Specifications

General IconGeneral
BrandKeysight Technologies
ModelMSO-X 2002A
CategoryTest Equipment
LanguageEnglish

Related product manuals