340 Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
25 UART/RS232 Triggering and Serial Decode
6 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode
data.
You can use the horizontal Zoom window for easier navigation of the acquired data.
See Also “Interpreting UART/RS232 Decode" on page 340
“UART/RS232 Totalizer" on page 341
“Interpreting UART/RS232 Lister Data" on page 342
“Searching for UART/RS232 Data in the Lister" on page 342
Interpreting UART/RS232 Decode
Angled waveforms show an active bus (inside a packet/frame).
Mid-level blue lines show an idle bus.
When using 5-8 bit formats, the decoded data is displayed in white (in binary,
hex, or ASCII).
NOTE
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow enough
that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
softkey to set the trigger mode from Auto to Normal.