EasyManuals Logo

Keysight E36200 Series User Manual

Keysight E36200 Series
249 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
If the output load resistance is less than the voltage setting divided by the current setting, the instrument will
operate in constant-current (CC) mode. The voltage will equal the current multiplied by the load resistance.
From the remote interface:
A channel parameter is required with each SCPI command to select an output. For example, (@1) selects output 1
and (@1,2) selects output 1 and 2. The output list must be preceded with an @ symbol and be enclosed in
parentheses ().
To set only output 1 to 5 V and 1 A:
APPL CH1 5, 1
To enable only output 1:
OUTP ON,(@1)
To enable output 1 and output 2:
OUTP ON,(@1,2)
To measure the output voltage and current of output 1:
MEAS:VOLT? (@1)
MEAS:CURR? (@1)
68
Keysight E36200 Series User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E36200 Series and is the answer not in the manual?

Keysight E36200 Series Specifications

General IconGeneral
BrandKeysight
ModelE36200 Series
CategoryPower Supply
LanguageEnglish

Related product manuals