Measurement
103
mm port,
connect a
test fixture,
and then
connect the
device to the
fixture
contacts
1
Although it has a cable, it falls into the category of direct
connection type test fixture for adapter setup.
The user calibration data and fixture compensation
data can be saved in the mass storage of the E4990A.
Other topics about Calibration
A. Direct Connection Type Test Fixture
This section describes the calibration procedures to follow when a direct
connection type test fixture is connected to the four-terminal pair port of
the E4990A and a device is connected to the test fixture contacts.