E4990A Help
490
SCPI.IEEE4882.WAI
INIT
:INIT<Ch>
Type
Command
Syntax
:INITiate<Ch>[:IMMediate]
Description
This command changes the state of each channel to the initiation state in
the trigger system.
When this command is executed for a channel in the idle state in the
trigger system, it goes into the initiation state immediately. Then, after
measurement is executed once, it goes back to the idle state.
If this command is executed for a channel that is not in the idle state or a
channel for which the continuous initiation mode is set to ON (setting by
which the trigger system initiates continuously) in the trigger system
, an
error occurs when executed and the object is ignored.
Examples
Ana.WriteString ":INIT", True
Related Commands
:INIT:CONT
Equivalent Softkey
Trigger > Single (When :INIT:CONT is OFF.)
COM Command Syntax (Internal Control Only)
SCPI.INITIATE(ch).IMMEDIATE
:INIT<Ch>:CONT
Type
Command/Query
Syntax
:INITiate<Ch>:CONTinuous {ON|OFF|1|0}
:INITiate<Ch>:CONTinuous?
Description
This command turns ON/OFF the continuous initiation mode (setting by
which the trigger system initiates continuously) in the trigger system.
Variable