Results of Analysis
Determine the Self-resonance Frequency and Resonant Impedance
Other topics about Measurement Example
Determine the Self-resonance Frequency and Resonant Impedance
1. Activate Trace 1 (you may use touch-screen or by toggling between
Trace Prev and Trace Next key).
2. Press Marker Search key to display the Marker Search menu.
3. Select Tracking > ON to turn on the search tracking function. When
this function is enabled, the instrument performs a search action
upon completion of each sweep cycle.
4. Select Min to move the main marker to the minimum |Z| value.
Since the search tracking function is enabled, the instrument detects
and moves the main marker to the minimum |Z| value each time it
has completed one sweep cycle.