Programming
265
Next i
Analyzer.WriteString LoadF(n), True
End Sub
Starting a Measurement (Trigger) and Detecting the Completion of a
Measurement
Starting a Measurement (Trigger) and Detecting the Completion of a Measurement
Trigger System
Starting a Measurement Cycle
Waiting for the End of Measurement
Detecting Occurrence of an Error
Trigger System
Overview
System-Wide States and Transitions
Channel-Wide States and Transitions
Other topics about Starting a Measurement (Trigger) and Detecting the
Completion
Overview
The trigger system is responsible for such tasks as detecting the start of a
measurement cycle (triggering) and enabling/disabling measurement on
each channel. As shown in the following figure, the trigger system has two
types of states: system-wide and channel-wide. The system-wide state can
be "Hold", "Waiting for Trigger", or "Measurement", while the channel-wide
state can be "Idle" or "Initiate".