206
Revision
format is A.xx.xx.
firmware update.
Option
Installed Options Number.
This is managed by the
license key.
The upgrade kit adds the
option.
OS ID
Windows License
Description
Immutable ID
IP
Address
IP address when the unit
is connected with LAN
It is depending on LAN
configuration.
MAC
Address
MAC Address of LAN
interface connection.
Immutable address unless
the CPU board is replaced
at repair.
USB ID
USB ID for VISA
connection
Immutable ID
Procedure
Follow the procedure below to get the product information.
1. Press System > Firmware Revision.
2. The Firmware Revision dialog is displayed.
3. Pressing PC Info shows the PC hardware information.
Backing Up License Key File
Backing Up License Key File
Other topics about Setting Control Functions
Backing Up License Key File
The license key for some options are kept in a text file (.lic) located at
Factory(E:)\LICENSE\. The E4990A refers to the license codes in all of .lic
files under the directory.
If there is no license code, an error occurs and the E4990A measurement
application works as simulator only.
When you change the storage, copy the .lic files as back up. You can re-
create the license number at http://www.agilent.com/find/softwarelicense