110
4. Perform the fixture compensation at the device contacts (electrodes)
of the fixture. See Fixture Compensation for the fixture compensation
procedure.
When an exclusive test fixture is connected to the 7-
mm port of the 42942A which is also connected to the four-
terminal pair port of the E4990A, port extension compensation
is not required.
Other topics about Calibration
E. 42942A, 7-mm Port Extension
Perform the following calibration when the 7-mm port of a 42942A
connected to the E4990A is extended with a one-port configuration and the
device is connected to the cable’s end with a minimal connection.