474
This command sets/gets the number of divisions in all the graphs.
Variable
Parameter Value
Description Number of divisions of graph
Data Type Numeric type (Integer)
Range 4 to 30
Preset Value 10
Resolution 2
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Long
Var= 20
Ana.WriteString ":DISP:WIND1:Y:DIV " & Str(Var), True
Ana.WriteString ":DISP:WIND1:Y:DIV?", True
Var=Ana.ReadNumber
Related Commands
:DISP:WIND:TRAC:Y:PDIV
:DISP:WIND:TRAC:Y:RLEV
:DISP:WIND:TRAC:Y:RPOS
Equivalent Softkey
Scale > Divisions
COM Command Syntax (Internal Control Only)
SCPI.DISPLAY.WINDOW(Ch).Y.SCALE.DIVISIONS
FORM
:FORM:BORD
Type
Command/Query
Syntax

Table of Contents

Other manuals for Keysight E4990A

Related product manuals