324
0 Not used Always 0
1
Trace 1
Limit Test
Fail
Set to "0" when a measurement cycle
begins;
set to "1" when the measurement cycle
finishes and returns "fail" as the limit test
result for trace 1.
2
Trace 2
Limit Test
Fail
Set to "0" when a measurement cycle
begins;
set to "1" when the measurement cycle
finishes and returns "fail" as the limit test
result for trace 2.
3
Trace 3
Limit Test
Fail
Set to "0" when a measurement cycle
begins;
set to "1" when the measurement cycle
finishes and returns "fail" as the limit test
result for trace 3.
4
Trace 4
Limit Test
Fail
Set to "0" when a measurement cycle
begins;
set to "1" when the measurement cycle
finishes and returns "fail" as the limit test
result for trace 4.
5 to 15 Not used Always 0
Issuing the *CLS command clears all the bits in the questionable limit
channel status event register.