Keysight E4990A - Specifying Measurement Conditions

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quick Start
43
2. Fasten two of the four BNC connectors to the corresponding test
connectors by gradually turning the BNC connectors' rotation levers
until each pair of connectors is securely connected. Be sure to align
the grooves on both sides.
3. Simultaneously turn the fixture's two fastening screws clockwise so
that the fixture is secured to the instrument.
4. Finally, secure the remaining two BNC connectors of the fixture by
turning their rotation levers clockwise.
Reverse the above procedure when removing the
16047E Test Fixture.
Turn ON the Power
Press the power switch to turn ON the power to the E4990A. The E4990A
performs a power-on self-test. During the self-test, the model name,
firmware revision number/date, options, copyright notice, and other
information appear on the LCD. When the self-test is completed, the
measurement screen appears on the LCD.
Specifying Measurement Conditions
Initialize the E4990A to the Preset State
Select Fixture
Select |Z|-Theta as the Measurement Parameter
Select Frequency in Log as the Sweep Parameter
Set the Sweep Start/Stop Value
Set the Measurement Speed
Other topics about Measurement Example
Next, you need to specify how your E4990A should perform the
measurement.