EasyManua.ls Logo

Keysight E4990A - Parameter Description

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E4990A Help
192
Data structure of the Touchstone file consists of a header part and a data
part. The contents of the file is text data, which is ready to be read with a
general text editor.
The header part consists of the returned value of *IDN?, file created date,
calibration state, list of all S parameters of a specified port, and format
information.
The header parts of s1p are shown below.
Header of s1p
!Agilent Technologies,E4990A,<ID>,<FW Revision>
!Date <Date>
!Data:
!Freq S11
# Hz S FMT R 50
The option line starting with "#" is placed at the beginning of the file. The
structure of the option line is as follows:
Parameter Description
Hz
Indicates the frequency unit. For the E4990A, it is fixed
to Hz.
S
Specifies the type of parameters in data lines. For the
E4990A, "S" is specified that indicates the S parameter.
FMT
Specifies the format. This can be changed by the
selection.
RI = Real number - imaginary number
MA = Linear magnitude - angle
DB = Log magnitude - angle
R 50
Specifies the characteristic impedance. For the E4990A
specified is "R 50", meaning 50 .
The following figures show the data structures of files saved in Touchstone
format.
1-port Touchstone file
Restrictions when saving data in Touchstone format