Programming
491
Description ON/OFF of the continuous initiation mode
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value
OFF
Query Response
{1|0}<newline><^END>
Examples
Dim Var as Long
Ana.WriteString ":INIT1:CONT ON", True
Ana.WriteString ":INIT1:CONT?", True
Var=Ana.ReadNumber
Related Commands
:INIT
Equivalent Softkey
Trigger > Hold | Continuous | Hold All Channels (This is not exact
equivalent softkey.)
COM Command Syntax (Internal Control Only)
SCPI.INITIATE(Ch).CONTINUOUS
LXI
:LXI:IDEN
Type
Command/Query
Syntax
:LXI:IDENtify[:STATe] {ON|OFF|1|0}
:LXI:IDENtify[:STATe]?
Description
This command sets or gets the LXI Status Indicator state.
Variable
Parameter Selection Option
Description ON/OFF LXI Control Identification