Measurement
105
This calibration procedure is applied to the following connections:
The 16048G (1-m Cable) is connected to the E4990A and device is
connected to the cable’s end with a test fixture.
The 16048G (1-m Cable) is connected to the E4990A and 16452A
(Liquid Test Fixture) is connected to the end of the cable.
The 16334A (Tweezer Type Test Fixture) is connected to the E4990A.
The 16451B (Dielectric Test Fixture) is connected to the E4990A.
1. Make sure that the adapter selection is set to 4TP 1M. Perform
Adapter Setup if it is not done yet.
2. Perform the fixture compensation in accordance with Fixture
Compensation.