EasyManua.ls Logo

Keysight E4990A - Page 427

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming
417
Range 1|2|3|4
Preset Value
2
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Long
Var= 4
Ana.WriteString ":CALC1:PAR:COUN " & Str(Var), True
Ana.WriteString ":CALC1:PAR:COUN?", True
Var=Ana.ReadNumber
Related Commands
:CALC:PAR:SEL
Equivalent Softkey
Display > Num Of Traces
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).PARAMETER.COUNT
:CALC<Ch>:PAR<Tr>:DEF
Type
Command/Query
Syntax
:CALCulate<Ch>:PARameter<Tr>:DEFine
{Z|Y|R|X|G|B|LS|LP|CS|CP|RS|RP|Q|D|TZ|TY|VAC|IAC|VDC|IDC|IMP|AD
M}
:CALCulate<Ch>:PARameter<Tr>:DEFine?
Description
This command sets/gets the measurement parameter. The VDC or IDC can
be selected only at bias sweep or log bias sweep.
Variable
Parameter Selection Option
Description Measurement parameter
Data Type Character string type (String)
Range
Z: Absolute impedance value
Y: Absolute admittance

Table of Contents

Other manuals for Keysight E4990A

Related product manuals